Storage and transport conditions
Temperature - electrodes from 0°C to +40°C
Temperature - device from -10°C to +50°C
Relative humidity less than 93% (non-condensing)
Atmospheric pressure from 700hPa to 1060hPa
Note: If you want to store the electrodes for more than a month,
keep them at temperatures between +5°C and +27°C.
Note: Do not store the electrodes in the freezer or refrigerator. Do not expose
them to high temperatures, nor immerse them in water, or leave them
outside of the plastic bag.
Explanation of symbols
Symbols on the adapter and the device
- This symbol indicates that the adapter is double insulated (Class II),
according to IEC 60601-1 (Fig. 5).
- This symbol means: Do not throw away with the normal household
waste. For further instructions, see chapter ‘Recycling’ (Fig. 2).
Symbols on the device
- This is the symbol for serial number. This symbol is followed by the
manufacturer’s serial number (Fig. 6).
- This is the symbol for reference number. This symbol is followed by
the manufacturer’s reference number, which is the type number of
the device (Fig. 7).
- This symbol on the device means: protected against access to
hazardous parts with a nger and against vertically falling water drops
when tilted up to 15 degrees (Fig. 8).
- This symbol means that the part of the device that comes into
physical contact with the user (also known as the applied part)
is of type BF (Body Floating) according to IEC 60601-1.
The applied parts are the electrodes (Fig. 9).